Laser ranging distomats
oct 2002
Testing TRelectronics distomat
(apr-jun11)
apr11: testing accurange
3000 distomat replacement.
1991:How much vertical motion do we
have with the DI1001 distomats at 157 meters.
azimuth to use when vertically aligning
the distomats
history
monitoring
the distomat dropouts and the modification log
Rfi
caused by the distomats
wild di1001 specs: ( I think this was one of the orignal distomats).
- accuracy: 5mm + 5ppm
- range: 1km (1prism)
- measurement time: 1,5 sec
- production year: 1989-2001
Azimuth angles to use
when aligning the distomats. (top)
The easiest way to align the distomats is to
position
the azimuth arm perpendicular to the corner where the distomat is
pointing.
This should leave that corner at the average platform height. It does
not
matter where the dome/ch is.
Azimuth angles:
- distomats 1,6 (corner T12) move the azimuth to 92.87,
272.87, or
452.87 degrees (whichever is closer)
- distomats 2,3 (corner T4) move the azimuth to 32.87, 212.87, or
392.87
degrees (whichever is closer).
- distomats 4,5 (corner T8) move the azimuth to 152.87,
332.87 or
512.87
degrees (whichever is closer).
Try and center the distomats vertical range on this position.
Some things to keep in mind.
- Make sure the tiedowns are tracking the average height of the
platform
before you start the adjustments. Look at the tiedown Status
display in the control room. The mode should be tieTrk for all 3
tiedowns.
If it does not say this, ask the operators to do a tietrk command
- The platform should be close to the reference position. Look on
the laser
ranging display (the plot of height versus hour of day to the
left of the observers workstation). On the upper right it has the time,
platform position, and the platform reference position. The time should
be within the last 3 or 4 minutes. The last position should be within
.05
feet of the reference position.
- If the software tries to take a measurement while you are
adjusting the
distomat, it may get a no reading. In that case it will switch and use
the temperature to position the tiedowns. This is probably not a
problem.
History
- 2010
- 20sep10: distomat 2 replaced with spare.
- 17sep10: distomat 2 has been failing a lot. when window removed
rcvd signal goes from 1 to 5. today replaced window with a window that
was on the kronos time system in building 1. rcv signal with window on
is now 2-2.5.. close to what other distomats have. (but distomat
continued to fail). Willy says there is an error registering on the
distomat.
- 20oct08: new reflectors at 2,3,4,5 installed
- 10oct08: new reflectors at 6,1 installed
- 10jan07: all dists window cleaned. dist2,3 were realigned
- 2007
- 24jan07: distomat 5 failed (sn 101115) with error 72. Replaced
with
spare
(sn 101036?).
- apr07: some distomats were covered by the containment for a few
days. Probably need to realign the distomats when painting done. For
apr07->jul07 data was used even when ch not in stow position. This
was needed since the painting did not leave the ch at stow.
- jul07: distomat 4 and 5 were off for a long time because of the
painting containment.
- aug,sept.. some distomats covered by painting containment.
- 2006
- 13nov06: dist 3( sn xxx) was replaced with s/n 10115) from leica.
The
external
window was replaced with a new one.
- 02oct06: dist5 window cleaned and realigned around 13:45
- 02oct06: dist6 (sn: 101115) was replaced by sn:10037. Cleaned and
then
realigned.
- 11sep06: all distomat windows cleaned with fuller anti-fog.
- 07sep06: morning. replaced distomat 5 (sn 100337) with
distomat
back
from maintenance (sn 101036).
- 06jul06: dist 4 (sn 101036) replaced by dist (sn 101075) and then
realigned.
sn 101036 will be sent back to fab for maintenance.
- 05jul06: all distomat windows cleaned.
- 21apr06: realigned D4. Looks like it got knocked out of alignment
by
the
previous days work. D5 was also realigned.
- 20apr06: d1 switched out for maintenance. It was replaced with
the
spare
that just came back from leica. windows cleaned and fans turned on for
distomats: 1,2,3,4 (5,6 will be done tomorrow).
- 17mar06:d1 window cleaned inside,out with anti fog spray 14:00.
- 16mar06:d2 window changed 11:25.
- 14mar06: d1 window changed, realigned. Used a different glue with
this
window :flexible silicone adhesive silver plate aluminum from chomerics.
- 06mar06: d1,d2,d3 exterior window cleaned with anti fog.
- 23feb06: distomat 2,3 outside window cleaned with EXXENE
antifog.
- 07feb06: new windows distomats: 3,4,5.
- 06feb06: new window distomat 6.
- 27jan06: afternoon all fans turned off. lenses cleaned.
- 19jan06: fan installed dist5
- 18jan06: fan installed dist 4
- 17jan06: fan installed dist 3
- 13jan06: fans installed dist 1 and 2.
- 11jan06: put fan in distomat 6 to try and improve temperature
circulation
inside canister.
- 2005
- 30may05 6pm distomat 4 stopped working. When examined, the heater
was
off.
Probably no power getting to distomat. On next day found that the
filter
for the ac line that is right inside where the acpower enters the
distomat
canister was toast...
- 31jan05:
all distomats realigned.
<-