Dewar temperatures
march,2002
The receiver systems have temperature sensors
that measure the physical temperature of the 16 Kelvin, 70 Kelvin, and
the OMT stages of the receiver. These temperatures are recorded daily.
They are plotted below by year and for the last 5 and 30 days (updated
each night). The color plots overplot all of the receivers on one plot
using color to differentiate between the receviers. The black and white
plots have a mosaic with 1 plot per receiver (if you don't have a color
plotter, then you can make hardcopy with the black and white version).
Some receivers do not have the temperature diodes to measure the 70K or
the OMT stages. These plots will appear blank. See using
idl to plot the dewar temperatures for routines that allow you to plot
this data interactively.
Notes:
2004
08mar04 - cbh now included in plots.
2003
13jan03- lbn dewar warmed up for maintenance. xband 16K stage drops by
5K since now 2 rather than 3 dewars on the compressor.
15jan03- lbn op-amp for tempmonitor conversion I to V added bypass capacitors.
Drifting should settle down..
16jan03- xb,sbn,lbn 1 compressor changed to (xb,sbh) (sbn,lbn)
26mar03 - xb,sbn,lbn back to 1 compresso sbh on 1 by itself.
04apr03 - lbn changed diode monitoring circuit to use old diode curve.
07apr03 - changed sbn diode board to use old diode curve..updated
software
2002
18feb02 xb - xband moved to share compressor with sbn,lbn. 16K stage jumped
a few degrees.
06apr02 lbw- left vacuum pump running on dewar. 70K 120->80K. must be a
leak in dewar.
27jun02 lbw - warmed up lbw, regreased seals, added vacion pump (but not
on yet), then cooled down.
10jul02 lbw - turned on vacion pump.
17sep02 lbn,xb - sbn no longer on compressor with xb,lbn (down for repairs).
xb,lbn temps decreased.
24sep02 lbn,xb,sbn back to being on same compressor.
20dec02 switched to use continous sampling. last5 plots sampled every 22
secs, last_30 plots smoothed to 4 minutes.
processing: x101/dwtemp/dwtemp.pro